Fluke 79 Series II LCD Repair Kit – How To


Display Repair Kit For Fluke 79 Series II

Compatible with: Fluke 79 Series II Multimeters

Download Instructions may be found here: Download Fluke 70 Series Instructions

Display Kits and Replacement Parts may be found at MultimeterParts.com

Equipment (Tools) Needed

  • Phillips Screwdriver (Not included)
  • Small Flathead Screwdriver, or equivalent (Not included)
  • Tweezers (Not included)
  • Alcohol Pad
  • Latex Gloves (Surgical Gloves), or similar
  • Pencil Eraser
  • (2) Elastomeric Connectors
    *Caution*
    ESD (Electrostatic Discharge) awareness and practices should be followed to minimize potential damage.
  • Discharge personal static prior to handling
  • Avoid plastic, vinyl, and Styrofoam in work area
  • Perform replacement at a static free work station

**Warning**

Remove All Input Signals and Test Leads Before Proceeding

Step 1: Case Disassembly

  1. Disconnect live source from the Fluke Meter’s input Turn the Fluke meter to Off position
  2. Using Phillips head screwdriver, remove the screws from the bottom (lower) Fluke case
  3. Separate the Top and Bottom Case

***Now that the casing is open and the inside is exposed, it is recommended to put on the Latex Gloves provided in the kit as an ESD Precaution***

Gloves are not needed if ESD Wrist straps are used.

Step 2: Replacing Elastomeric Connectors

  1. Use the small Flathead Screwdriver to remove the LCD Faceplate

    ***Warning: Be careful! Plastic becomes brittle with age. Insert the small flathead screwdriver into the opening and very slightly pry up.
  2. Remove the Faceplate to expose the LCD
  3. Use your fingers to remove the LCD and Elastomers. You can use your small Flat Head Screwdriver to carefully pry one side of the LCD to loosen it up. Please Use Caution.
  4. Carefully Remove the LCD and Elastomers from PCB Board.
    • LCD may stick a little. It’s normal. Carefully lift from the edges until LCD Comes out.
    • Elastomeric connectors may, come out along with the LCD. Again, this is normal.
    • Use tweezers to remove Elastomeric Connectors if necessary.
  5. Insert the Two New Elastomers
  6. Clean the LCD Contacts with Eraser and Alcohol Pads
    FRONT BACK
    • Use the Pencil Eraser to rub along the Top and Bottom Edges of the LCD then use the Alcohol Pads to wipe clean around the edges of the LCD

Step 3: Re-Assemble Multimeter

  1. Place the clean LCD over the Elastomeric Connectors. Make sure the LCD Digit Side is face up and not upside down.
  2. Place the LCD Faceplate over the LCD and gently press down and snap into place
  3. Seat the Top Case and Bottom Case back together
  4. Insert the Screws to the Bottom Case and use Phillips Screwdriver to tighten the cases together.

Enjoy your nice digital display.

***IMPORTANT***

If your display is physically damaged, or cracked, or if the LCD Cover is broken or cracked, then this procedure will not work. The LCD cover, or LCD would have to be replaced separately to fix the display.

LCD Repair Kit

We've been serving the test and measure industry for over 20 years, and know how important it is to have clean and clear display readings. We look forward to accommodating your LCD Repair Kit needs.

Recent Posts